5-56
TOS9213S/ TOS9213AS
5.11.4 Judgement on the program
■
PASS judgement
When the interval time for each step is set to other than HOLD, the program makes
a PASS judgement if a PASS judgement is made for every step (only if the pro-
grammed repeat test is set to END).
A PASS judgement is made after the interval time for the previous step has elapsed.
The tester then returns to the READY status.
If the interval time for any step is set to HOLD, the test does not proceed to the next
step unless the START switch is pressed.
■
FAIL judgement
Once a FAIL judgement is made while the program is being run, the test stops at
that step.
Check the failed step and press the STOP switch.
If the START switch is pressed again, the program starts from the beginning.
5.11.5 Exiting the program
To exit the PROGRAM mode in order to return to the normal mode, press the DCW
key or IR key.
The LED on the AUTO key goes off, and the test-condition setting screen appears.