6-14
TOS9213S/ TOS9213AS
6.4
STATUS OUT Connector
The STATUS OUT connector is an output connector to be connected to the warning
light.
Choose warning statuses on the System screen, from among H.
V
ON, TEST, PASS,
READY, U FAIL, L FAIL, CONTACT FAIL, PROTECTION, and POWER ON.
If two or more statuses are chosen at a time, the logical sum of them, is outputted.
For settings on the System screen, see "5.8 System Settings".
Maximum output voltage: 24
V
DC; maximum output current: 100 mA
Compatible cables:
• Solid wire, ø0.65 (AWG22)
• Stranded wire, 0.32 mm
2
(AWG22); element wire diameter: ø0.18 or more
Connecting to the connector
1.
Remo
v
e approximately 11 mm of coating from the tip of the signal
ca
b
le.
2.
Insert a scre
w
dri
v
er into A, and open B.
3.
Insert the ca
b
le into B. Take care to pre
v
ent the coating from
b
ecoming
ca
u
ght in B.
4.
Gently p
u
ll on the ca
b
le to confirm that it is sec
u
rely connected.
•
Do not extract a c
u
rrent exceeding the maxim
u
m rated c
u
rrent of 100 mA
from the +24
V
internal po
w
er s
u
pply. Once a c
u
rrent exceeds the maxi-
m
u
m rated c
u
rrent, the o
v
erc
u
rrent protection f
u
nction acti
v
ates and “24
V
OCP” flashes on the LCD.
The tester enters the PROTECTIO
N
stat
u
s. If the o
v
erc
u
rrent protection
f
u
nction acti
v
ates, t
u
rn off the PO
W
ER s
w
itch to pre
v
ent an o
v
erc
u
rrent.
If an o
v
erc
u
rrent flo
w
s for an extended period, the internal circ
u
it may mal-
f
u
nction.
A
B
Shank diameter: ø3; tip width: 2.6 mm
11 mm
CAUTION