![Kikusui TOS9213AS Operation Manual Download Page 169](http://html.mh-extra.com/html/kikusui/tos9213as/tos9213as_operation-manual_1966734169.webp)
TOS9213S/ TOS9213AS
7-45
DCW:STORE (D:STOR)
Stores the current settings for a DC withstanding voltage test. Note that the memory
name is not stored (the existing memory name is not replaced).
Message
Setting:
DCW:STORE_<
memory number
>
D:STOR_<
memory number
>
Program data
Data format: Integer
Set value:
0 to 99
Resolution:
1
(Example) To store the current settings for a DC withstanding voltage test to mem-
ory 10,
D:STOR 10
IR:STORE (I:STOR)
Stores the current settings for an insulation resistance test. Note that the memory
name is not stored (the existing memory name is not replaced).
Message
Setting:
IR:STORE_<
memory number
>
I:STOR_<
memory number
>
Program data
Data format: Integer
Set value:
0 to 99
Resolution:
1
(Example) To store the current settings for an insulation resistance test to memory
10,
I:STOR 10