7-32
TOS9213S/ TOS9213AS
IR:SCANW (I:SCANW)
Sets HIGH/LOW/OPEN for 16 scanner channels at a time using 16 bits for an insu-
lation resistance test. Also inquires about the settings for all 16 channels.
Message
Setting:
IR:SCANW_<
setting 1
,
setting 2
>
I:SCANW_<
setting 1
,
setting 2
>
Query:
IR:SCANW?
I:SCANW?
Program data
<setting 1>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
LOW is represented by 0 and HIGH by 1.
Program data
<setting 2>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
OPEN is represented by 1 and NOT OPEN by 0.
(Example) To set scanner channels 1 through 4 to LOW, scanner channels 5
through 8 to HIGH, and the other channels to OPEN,
I:SCANW #H00F0,#HFF00
OPEN in setting 2 has priority over setting 1.
Response
To I:SCANW
?
, the settings for all 16 channels of the high-voltage scanner are
returned as “setting 1” and “setting 2” using 16 bits.
(Example) If scanner channels 1 through 5 are currently set to HIGH, channels 6
through 10 to OPEN, and the other channels to LOW,
31,992
is returned.
IR:CONTACTCHECK (I:CCH)
Sets ON/OFF of the contact check function for an insulation resistance test. Also
inquires about the status of the function.
Message
Setting:
IR:CONTACTCHECK_<{ON|OFF}>
I:CCH_<{ON|OFF}>
Query:
IR:CONTACTCHECK?
I:CCH?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the contact check to ON,
I:CCH ON
Response
To I:CCH
?
, the current setting for the contact check is returned.