10.1 Withstanding Voltage Test Mode
*
1 Limitation on output
The tester’s withstanding voltage generator is designed to radiate half as much heat as the rated output, in consideration of
the size, weight, cost, and other factors of the tester. It is therefore necessary to use the tester within the ranges specified
below. Operations deviating from these ranges may heat the output section excessively, thereby activating the protective
circuit. In such a case, suspend the test and wait until the temperature falls to the normal level.
O
u
tp
u
t limitation in
w
ithstanding
v
oltage testing
(Output time = voltage rise time + test time + voltage fall time)
*
2 About the discharge time setting
If you set the discharge time to 0.0 s or if the voltage between the output terminals exceeds approximately 30
V
even after
the specified discharge time has passed, the TOS9213S/ TOS9213AS will continue discharging until the voltage between
the output terminals falls below approximately 30
V
.
O
u
tp
u
t section
DC
O
u
tp
u
t-
v
oltage range
0.05 k
V
to 10.0 k
V
DC
Resol
u
tion
10
V
Acc
u
racy
±(1.5 % of s 20
V
)
Maxim
u
m rated load
(*1)
50
W
(10 k
V
/5 mA)
Maxim
u
m rated c
u
rrent
5 mA
Ripple
N
o load at 10
k
V
100
V
p-p Typ.
Maxim
u
m rated
load
100
V
p-p Typ.
V
oltage reg
u
lation
1 % or less [maxim
u
m rated load
→
no load]
Short-circ
u
it c
u
rrent
40 mA Typ.
Discharge f
u
nction
Forced discharge at the end of test (discharge resistance: 500 k
Ω
)
The discharge time can
b
e set to a
v
al
u
e from 0.5 s to 300 s (*2).
Start
v
oltage
The
v
oltage at the start of the test can
b
e set as the start
v
oltage.
Setting range
0% to 99% of the test
v
oltage (resol
u
tion of 1%)
O
u
tp
u
t-
v
oltage monitoring
f
u
nction
If the o
u
tp
u
t
v
oltage exceeds ±(10% of s 50
V
), o
u
tp
u
t is c
u
t off and the
protection f
u
nction acti
v
ates.
V
oltmeter
Analog
Scale
10 k
V
AC/DC F.S
Acc
u
racy
±5% F.S
Indicator
Mean-
v
al
u
e responsi
v
e
Digital
Meas
u
rement
range
0.0 k
V
to 10.5 k
V
DC
Resol
u
tion
10
V
Acc
u
racy
±(1.0% of r 20
V
)
Response Mean-
v
al
u
e responsi
v
e (response time of 200 ms)
HOLD f
u
nction
The
v
oltage meas
u
red at the end of test is held d
u
ring the PASS and FAIL
period.
Am
b
ient
temperat
u
re
Upper reference
Pa
u
se
O
u
tp
u
t time
t
≤
40 °C
DC
2.5 mA < i
At least as long as the o
u
tp
u
t time
Maxim
u
m of 1 min
u
te
i
≤
2.5 mA
At least as long as the j
u
dgement
w
ait time (
W
AIT TIME)
Contin
u
o
u
s o
u
tp
u
t possi
b
le