10-4
TOS9213S/ TOS9213AS
*
3 When the GND LOW/GUARD setting is set to LOW, the humidity must not exceed 70 % rh.
*
4 With the GND set to GUARD, current measurement is disabled if the part of the DUT connected to the LOW terminal is
grounded, which poses extreme danger. Never ground the DUT. In ordinary operation, set the GND to LOW.
*
5 In the MID and SLOW modes, depending on the discharge method, the voltage monitoring function may operate and the
TOS9213S/ TOS9213AS may enter the PROTECTION status before UPPER FAIL detection takes place.
Time
Setting range for the
v
oltage
rise time (RISE TIME)
0.1 s to 200 s
Setting range for the test time
(TEST TIME)
0.3 s to 999 s (
W
ith the TIMER OFF f
u
nction)
Acc
u
racy
± (100 ppm + 20 ms)
Setting range for the j
u
dge-
ment
w
ait time (
W
AIT TIME)
0.3 s to 30 s [(RISE TIME + TEST TIME) >
W
AIT TIME]