TOS9213S/ TOS9213AS
7-17
Resolution:
0.01 E-3 for 1.00 E-3 to 5.50 E-3, 0.001 E-3 for 0.001 E-3 to
0.999 E-3
Unit:
A
Program data
<{ON|OFF}>
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the lower reference to 1 mA,
D:LOW 1E-3,ON
Response
To D:LOW
?
, the currently set lower reference and the ON/OFF status of the lower
judgment function are returned as “lower reference, ON/OFF.”
(Example) If the lower reference is currently set to 2.0 mA and the lower judgment
function to OFF,
“
2.00E-3,0
” is returned.
DCW:UPPER (D:UPP)
Sets the upper reference (UPPER) for a DC withstanding voltage test. Also inquires
about the set or setting upper reference.
• If the
u
pper reference is set a
b
o
v
e 5 mA, the protecti
v
e circ
u
it may
b
e acti-
v
ated. To pre
v
ent this, set the o
u
tp
u
t time
b
elo
w
1 min
u
te and the pa
u
se to
longer than the o
u
tp
u
t time.
Message
Setting:
DCW:UPPER_<
upper reference
>
D:UPP_<
upper reference
>
Query:
DCW:UPPER?
D:UPP?
Program data
Data format: Real number
Set value:
0.01 E-3 to 5.50 E-3
Resolution:
0.01 E-3 for 1.00 E-3 to 5.50 E-3, 0.001 E-3 for 0.001 E-3 to
0.999 E-3
Unit:
A
(Example) To set the upper reference to 5 mA,
A:UPP 5E-3
Response
To D:UPP
?
, the preset upper reference is returned.
(Example) If the upper reference is currently set to 2.00 mA,
2.00E-3
is returned.
CAUTION