5-6
TOS9213S/ TOS9213AS
5.4.1
Settings on the DCW1 screen
Setting the test voltage for DC withstanding voltage testing
The test voltage to be applied to a DUT can be set to 0.00 k
V
through 10.5 k
V
DC
(at a resolution of 0.01 k
V
).
To make settings, use the rotary knob with the cursor at the test voltage.
1.
To
b
ring the c
u
rsor to the test
v
oltage, press the F1 (
V
OLT) key (the
keys can also
b
e
u
sed).
2.
Using the
keys, mo
v
e the c
u
rsor to the target digit.
3.
Using the rotary kno
b
, set a test
v
oltage.
• When settings are being made, if the test voltage multiplied by the upper reference
exceeds 55 W, “READY” disappears and “O
V
ER 55 W” flashes at the top right of
the LCD to indicate that testing cannot be performed. In such a case, reduce the
test voltage or the upper reference.
Turning the lower judgement function ON/OFF
The lower judgement function can be turned on/off.
When the lower judgement function is on, the test ends with a FAIL judgement if
the measured current drops below the lower reference. The lower reference is
explained in the following section.
You can set the lower judgement function to ON/OFF using the SHIFT + F2 keys,
regardless of the location of the cursor. Each time the SHIFT + F2 keys are pressed,
ON and OFF alternate.
The
keys can also be used to move the cursor to ON/OFF of the lower
judgement function. If the cursor is at ON/OFF of the lower judgement function, the
rotary knob can be used to make settings.
VOLT LOWER UPPER
TIMER
1.00
k V
5.00
mA
3.0
s
DCW1
DCW1
LOWER
OFF
TIMER
OFF
READY
UPPER
−
F1 F2 F3 F4 F5
SHIFT
NOTE
1.00
k V
5.00
mA
3.0
s
DCW1
DCW1
LOWER
OFF
TIMER
OFF
READY
UPPER
−
F1 F2 F3 F4 F5
SHIFT
LOWER
ON/OFF
TIMER
ON/OFF