TOS9213S/ TOS9213AS
5-47
5.8.3
SYSTEM3
Setting the STATUS SIGNAL OUTPUT
Make settings to specify the timing of the output of a DC voltage of 24
V
from the
STATUS OUT connector on the rear panel.
To make settings for the STATUS SIGNAL OUTPUT terminal, use the function
keys or the rotary knob.
Each time a function key or the SHIFT + function keys are pressed, the item shown
on the key is turned on/off.
The
mark appears to the left of the selected item. Using the rotary knob, the items
shown with the mark on the left can be set.
Turn the rotary knob clockwise
: ON
Turn the rotary knob counterclockwise : OFF
Items set to ON are highlighted.
H.V ON
When H.
V
ON is set to ON, 24
V
DC is output during the test, while a
voltage remains between the output terminals, or during automatic test-
ing.
TEST
When TEST is set to ON, 24
V
DC is output during the test after the test
voltage reaches the preset value. That is, no voltage is output during the
RISE time or FALL time.
PASS
When PASS is set to ON, 24
V
DC is output during a PASS judgement.
READY
When READY is set to ON, 24
V
DC is output while READY is dis-
played on the LCD.
U FAIL
When U FAIL is set to ON, 24
V
DC is output while FAIL is output in an
upper judgement.
L FAIL
When L FAIL is set to ON, 24
V
DC is output while FAIL is output in a
lower judgement.
C FAIL
When C FAIL is set to ON, 24
V
DC is output while FAIL is output in a
contact check.
H.V
ON
TEST
PASS
READY
SYSTEM 3
SYSTEM 3
STATUS SIGNAL
OUTPUT :
F1 F2 F3 F4 F5
OFF ON
H.V ON
TEST
PASS
READY
OFF ON
U FAIL
L FAIL
C FAIL
PROTECT
POWER ON
SYSTEM 3
SYSTEM 3
STATUS SIGNAL
OUTPUT :
F1 F2 F3 F4 F5
OFF ON
H.V ON
TEST
PASS
READY
OFF ON
U FAIL
L FAIL
C FAIL
PROTECT
POWER ON
UPPER
FAIL
LOWER
FAIL
CONTACT
FAIL
PROTEC-
TION
POWER
ON
SHIFT