TOS9213S/ TOS9213AS
5-5
5.4
Settings for DC Withstanding Voltage
Testing
To make settings for a DC withstanding voltage test, use the DC Withstanding
V
olt-
age-Test Setting screen (DCW).
To jump to this screen (DCW1), press the DCW key. The LED on the DCW key
lights up.
The DC Withstanding
V
oltage-Test Setting screen has three pages, from DCW1 to
DCW3. To move between these pages, press the SHIFT key + keys. To return
to DCW1 from DCW2 or DCW3, press the DCW key.
• No setting can be made in the KEYLOCK status.
The three DCW pages allow the following settings to be made:
DCW1
• Test voltage
• Lower reference (LOWER) and ON/OFF of the lower judgement function
• Upper reference (UPPER)
• Test time (TEST TIME) and ON/OFF of the timer function
DCW2
• Start voltage
•
V
oltage rise time (RISE TIME)
• Judgement wait time (WAIT TIME)
• Discharge time (DISCH TIME)
• LOW/GUARD settings for the GND
DCW3
• Channel settings for the high-voltage scanner
• ON/OFF of the contact check function
To move the cursor between items, use the
keys.
When a function is shown on the F1 to F5 keys, they can be used to jump to the tar-
get item. To make settings for items displayed on these keys, press the SHIFT key +
F1 to F5 keys.
NOTE