7-26
TOS9213S/ TOS9213AS
IR:UPPER (I:UPP)
Sets the upper reference (UPPER) and ON/OFF of the upper judgment function for
an insulation resistance test. Also inquires about the set or setting upper reference
and the ON/OFF status of the upper judgment function.
Message
Setting:
IR:UPPER_<
upper reference
, {ON|OFF}>
I:UPP_<
upper reference
, {ON|OFF}>
Query:
IR:UPPER?
I:UPP?
Program data
Resistance
v
al
u
e-
b
ased j
u
dgment
Data format: Real number
Set value:
0.01 E6 to 9.99 E9
Resolution:
0.01 E6 (0.01 E6 to 9.99 E6), 0.1 E6 (10.0 E6 to 99.9 E6)
1 E6 (100 E6 to 999 E6), 0.01 E9 (1.00 E9 to 9.99 E9)
Unit:
Ω
Program data
C
u
rrent
v
al
u
e-
b
ased j
u
dgment
Data format: Real number
Set value:
0.01 E-6 to 1.00 E-3
Resolution:
0.1 E-6 (0.1 E-6 to 99.9 E-6), 1 E-6 (100 E-6 to 1000 E-6)
Unit:
A
Program data
<{ON|OFF}>
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the upper reference to 1.00 G
Ω
,
I: UPP 1.00E9,1
Response
To I:UPP
?
, the preset upper reference and the ON/OFF status of the upper judgment
function are returned as “upper reference, 1/0.”
(Example) If the upper reference is currently set to 5.00 G
Ω
and the upper judg-
ment function to OFF,
“
5.00E9,0
” is returned.