TOS9213S/ TOS9213AS
7-47
Program data
<
timer
{ON|OFF}
>
Data format: Character
Set value:
OFF, ON (0, 1)
Program data
<
start voltage
>
Data format: Integer
Set value:
0 to 99
Resolution:
1
Unit:
%
Program data
<
voltage rise time
>
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
Program data
<
WAIT TIME
>
Data format: Real number
Set value:
0.3 to 30.0
Resolution:
0.1
Unit:
s
Program data
<
GND
{GUARD|LOW}
>
Data format: Character
Set value:
LOW (0), GUARD (1)
Program data
<
contact check
{ON|OFF}
>
Data format: Character
Set value:
OFF, ON (0, 1)
Program data
<scanner Hi/Low
>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
LOW is represented by 0 and HIGH by 1.
Program data
<
scanner Open
>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
OPEN is represented by 1 and NOT OPEN by 0.
OPEN has priority over HIGH/LOW.
Response
To D:MEM
?
_
<
memory number
>, the contents of the memory specified by
memory number are returned. The memory contents are returned in the same order
as with program messages.
(Example) If memory 12, at inquiry, contains the memory name DCW12, test volt-
age of 5.50 k
V
, lower reference of 5.00 mA, upper reference of 5.00 mA,
test time of 2.0 s., lower judgment set to ON, timer set to ON, start volt-
age of 20%, voltage rise time of 0.2 s, WAIT TIME of 0.5 s, GND set to
LOW, contact check set to ON, scanners 1 through 5 set to Low, scan-
ners 6 through 10 set to High, and the other scanners set to Open, then,
“
DCW12,5.50E3,5.00E-3,5.00E-3,2.0,1,1,20,0.2,
0.5,0,1,992,64512
” is returned.