TOS9213S/ TOS9213AS
8-5
9.
Using the key, mo
v
e the c
u
rsor to the memory n
u
m
b
er.
10.
Using the rotary kno
b
, set the TOS6200’s memory n
u
m
b
er to
b
e
u
sed.
11.
Using the key, mo
v
e the c
u
rsor to the inter
v
al time to the right of the
memory n
u
m
b
er.
12.
Using the rotary kno
b
, set the inter
v
al time (0.2 s to 9.9 s, HOLD).
With the interval time set to HOLD, when the START switch is pressed, the
next step starts if the specified step is in the HOLD status.
13.
In accordance
w
ith the a
b
o
v
e proced
u
re, make settings for an earth
contin
u
ity test in the program step.
Upon completion of settings, press the AUTO key to return to the Program
screen.
The Program screen displays the total number of steps thus set. The tester
enters the READY status.
AUTO
AUTO
READY
PRG
00 : --UNTITLED--
TOTAL:10
01 EC 00-1.0s
02 IR 02-1.0s
03 EC 03-1.0s
Program name
Step No.
Program No.
Test step
Total number of steps
Repeat settings