TOS9213S/ TOS9213AS
7-33
(Example) If the contact check is currently set to ON,
1
is returned.
7.4.4
Messages common to all tests
This section explains the messages commonly used in all tests.
VDATA (VDAT)
Inquires about the monitor voltage for each test.
During the test, the real-time voltage is returned. After the test, the previous test
voltage is returned.
Message
Query:
VDATA?
VDAT?
Response
To
V
DAT
?
, the monitor voltage is returned for DCW in a DC withstanding voltage
test, and for IR in an insulation resistance test.
(Example) If the current monitor voltage in an DC withstanding voltage test is 1.5
k
V
,
1.50E3
is returned.
• In insulation resistance testing with the lower judgment function off, even if set-
tings are made so that the test voltage divided by the lower reference exceeds 1.1
mA, testing can be performed. However, when the voltage value which is out of
the range of ±(10 % of s50
V
) is detected during the test, the measurement
voltage value flashes on the LCD. In this case, -1 is returned.
IDATA (IDAT)
Inquires about the monitor current.
During the test, the measured normal value or the maximum value is returned,
depending on the MEAS MODE setting in the system settings. After the test, the
previous test current is returned.
Message
Query:
IDATA?
IDAT?
Response
To IDAT
?
, the monitor current is returned for DCW in a DC withstanding voltage
test.
(Example) If the present monitor current in an DC withstanding voltage test is 10.0
mA,
10.0E-3
is returned.
NOTE