7-24
TOS9213S/ TOS9213AS
7.4.3
Messages used exclusively for insulation resistance
testing
This section explains the messages used to check the test conditions and settings for
insulation resistance testing.
IR:TESTV (I:TES)
Sets the test voltage for an insulation resistance test. Also inquires about the preset
test voltage.
• This message is valid even during a test. However, make only fine voltage adjust-
ments rather than a large change to the voltage. A substantial change may activate
the protective circuit to shift the tester to the protection status.
Message
Setting:
IR:TESTV_<
voltage
>
I:TES_<
voltage
>
Query:
IR:TESTV?
I:TES?
Program data
Data format: Integer
Set value:
10 to 1020 (TOS9213S)
10 to 1530 (TOS9213AS)
Resolution:
1
Unit:
V
(Example) To set the test voltage to 500
V
,
I:TES 500
Response
To I:TES
?
, the currently set test voltage is returned.
(Example) If the test voltage is currently set to 250
V
,
250
is returned.
NOTE