TOS9213S/ TOS9213AS
8-7
8.3
Test Judgement
The tester reads the test results of the TOS6200 and displays a judgement on-screen.
FAIL judgement by the TOS6200
When the tester or TOS6200 makes a FAIL judgement at any step of the execution
of a program, the program stops at that step regardless of whether the repeat settings
of the tester program are made to RETURN or END, and regardless of the preset
interval time.
To cancel the FAIL status of a FAIL judgement in an earth continuity test, use the
STOP switch on the tester.
PASS judgement by the TOS6200
With the interval time for each step set to any value other than HOLD, if a PASS
judgement is made for all steps, a PASS judgement is made for the program as a
whole. (This applies only to the repeat settings made to END.)
A PASS judgement is made after the interval time has elapsed for the last step. The
tester then returns to the READY status.
If the interval time is set to HOLD in any step, it is not possible to proceed to the
next step unless the START switch on the tester is pressed.
To suspend the test, press the STOP switch on the tester. The tester returns to the
READY status.
• During the execution of a program, no buzzer may sound even if the TOS6200
makes a PASS judgement. This does not indicate any abnormality.
Displaying a communication error
In the following cases, “COM ERR” flashes and the PROTECTION LED lights up
to indicate a communication error:
• The POWER switch on the TOS6200 is turned off.
• The interface settings for the TOS6200 are not made as instructed in "8.1.2
• The RS-232C cable is disconnected or broken.
• At the start of a test, the TOS6200 is not in the READY status.
To cancel a communication error, press the STOP switch. Eliminate the cause of the
communication error, and restart the test.