3
Calibration Procedures
52
Keysight E3631A Service Guide
Test considerations
To ensure proper instrument operation, verify that you have selected the correct
power-line voltage prior to attempting any test procedure in this chapter. See
page 33 in chapter 2 for more information.
– Ensure the calibration ambient temperature is stable and between 20°C
and 30°C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables and test leads as short as possible, consistent with the
impedance requirements.
– Use shielded cables to connect the monitoring device such as oscilloscope
or RMS voltmeter to the power supply.
– Ensure all the devices are firmly connected.The alligator or crocodile clip is
not recommended for use in device connection.
Performance verification tests
The performance verification tests use the power supply's specifications listed in
Chapter 1,
You can perform two different levels of performance verification tests:
– Self-Test A series of internal verification tests that provide high confidence
that the power supply is operational.
– Performance Verification Tests These tests can be used to verify the power
supply's specifications following repairs to specific circuits. The
performance test procedures must be performed on each output.
CAUTION
The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the
power supply.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...