3
Calibration Procedures
62
Keysight E3631A Service Guide
7
Disconnect the oscilloscope and connect a RMS voltmeter in its place as
shown in
. The RMS voltage reading does not exceed the RMS limit
of 0.35 mV for each of the three outputs.
8
Repeat steps (1) through (7) for the remaining outputs.
Figure 3-3
Peak-to-peak ripple/noise measurement setup