Supplemental Characteristics
Table 1-2
Supplemental Characteristics
Item
Specifications
Output programming range (maximum programmable values)
+6V output
+25V output
–25V output
Voltage
0 to 6.18 V
0 to 25.75 V
0 to –25.75 V
Current
0 to 5.15 A
0 to 1.03 A
0 to 1.03 A
Temperature coefficient, ±(% of offset)
Maximum change in output/readback per °C after a 30-minute warm-up
+6V output
+25V output
–25V output
Voltage
0.01% + 2 mV
0.01% + 3 mV
0.01% + 3 mV
Current
0.02% + 3 mA
0.02% + 0.5 mA
0.02% + 0.5 mA
Stability, ±(% of offset)
Following a 30-minute warm-up, with the output in the ON state according to the operating mode (CC with load or CV), and
with a change in the output over 8 hours under constant load, line, and ambient temperature
+6V output
+25V output
–25V output
Voltage
0.03% + 1 mV
0.02% + 2 mV
0.02% + 2 mV
Current
0.1% + 3 mA
0.05% + 1 mA
0.05% + 1 mA
Voltage programming speed
Maximum time required for output voltage to settle within 1% of its total excursion (for resistive load). Excludes command
processing time.
+6V output
+25V output
–25V output
Full load up
11 msec
50 msec
50 msec
Full load down
13 msec
45 msec
45 msec
No load up
10 msec
20 msec
20 msec
No load down
200 msec
400 msec
400 msec
Isolation
The 0-6V supply is isolated from the ±25V supply up to ±240 Vdc. Maximum isolation voltage from any terminal to chassis
ground ±240 Vdc.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...