Specifications
1
Keysight E3631A Service Guide
27
[1]
Accuracy specifications are after an 1-hour warm-up and calibration at 25°C.
Voltage
<0.01% + 2 mV
Current
<0.01% + 250
m
A
Programming resolution
+6V output
+25V output
–25V output
Voltage
0.5 mV
1.5 mV
1.5 mV
Current
0.5 mA
0.1 mA
0.1 mA
Readback resolution
+6V output
+25V output
–25V output
Voltage
0.5 mV
1.5 mV
1.5 mV
Current
0.5 mA
0.1 mA
0.1 mA
Meter resolution
+6V output
+25V output
–25V output
Voltage
1 mV
10 mV
10 mV
Current
1 mA
1 mA
1 mA
Transient response time
Less than 50
m
sec for output recover to within 15 mV following a change in output current from full load to half load or vice
versa
Command processing time
Programming Commands : Maximum time for output to change after receipt of APPLy and SOURce commands : <50 msec
Readback Command : Maximum time to readback output by MEASure? command : <100 msec
The Other Commands : < 50 msec
Tracking accuracy
The ±25V outputs track each other within ± (0.2% of 20 mV).
Item
Specifications
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...