Calibration Procedures
3
Keysight E3631A Service Guide
71
7
Repeat steps (1) through (6) for the remaining outputs.
Normal mode current noise (CC ripple and noise)
The normal mode current noise is specified as the rms output current in a
frequency range 20 Hz to 20 MHz with the power supply in constant current
operation.
1
Turn off the power supply and connect the output to be tested as shown in
with a load resistor (1.2? for +6V supply and 25? for ±25V supplies)
across output terminals to be tested. Connect a rms voltmeter across the load
resistor. Use only a resistive load for this test.
2
Turn on the power supply and select the output to be tested using the meter
selection key on the front panel. Enable the outputs and set the display to the
limit mode. When the display is in the limit mode, program the current to
full-scale value and the voltage to the maximum programmable value.
3
The output current should be at the full-scale rating with the CC annunciator
on. If not lit, adjust the load so that the output voltage drops slightly until the
CC annunciator lights.
4
Divide the reading on the rms voltmeter by the load resistance to obtain rms
current. The readings should be within the limit specified below for each
output tested.
5
Repeat steps (1) through (4) for the remaining outputs.
Output
Specification
+6V Output
2 mA
+25V Output
0.5 mA
-25V Output
0.5 mA
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...