Service
5
Keysight E3631A Service Guide
117
631
System DAC test failed This test checks if the DAC hardware is functional. The
main controller U17 (U10 for serial MY53xx6xxx) sends a reference voltage data to
DAC and converts the DAC output to digital data to see if the digital data is within a
valid range.
632
P6V hardware test failed This test checks the the status of voltage and current error
amplifiers for +6V power circuit. If both amplifiers are not operational, the power
supply will beep and the error annunciator will be on.
633
P25V hardware test failed This test checks the the status of voltage and current
error amplifiers for +25V power circuit. If both amplifiers are not operational, the
power supply will beep and the error annunciator will be on.
634
N25V hardware test failed This test checks the the status of voltage and current
error amplifiers for -25V power circuit. If both amplifiers are n
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...