3
Calibration Procedures
74
Keysight E3631A Service Guide
CC Performance Test Record
Test Description
Actual Result
Specifications
Upper Limit
Lower Limit
CC Programming Accuracy @ 0 amps (I
O
)
+6V supply
+25V supply
-25V supply
+0.0100 A
+0.0040 A
+0.0040 A
-0.0100 A
-0.0040 A
-0.0040 A
CC Readback Accuracy @ 0 amps
+6V supply
+25V supply
-25V supply
I
O
+ 0.0100 A
I
O
+ 0.0040 A
I
O
+ 0.0040 A
I
O
- 0.0100 A
I
O
- 0.0040 A
I
O
- 0.0040 A
CC Programming Accuracy @ Full Scale (I
O
)
+6V supply
+25V supply
-25V supply
5.0200 A
1.0055 A
1.0055 A
4.9800 A
0.9945 A
0.9945 A
CC Readback Accuracy @ Full Scale
+6V supply
+25V supply
-25V supply
I
O
+ 0.0200 A
I
O
+ 0.0055 A
I
O
+ 0.0055 A
I
O
- 0.0200 A
I
O
- 0.0055 A
I
O
- 0.0055 A
CC Load Regulation
+6V supply
+25V supply
-25V supply
Maximum change:
<0.75 mA
<0.35 mA
<0.35 mA
CC Line Regulation
+6V supply
+25V supply
-25V supply
Maximum change:
<0.75 mA
<0.35 mA
<0.35 mA
CC Ripple/Noise
+6V supply
+25V supply
-25V supply
<2 mA
<0.5 mA
<0.5 mA
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...