3
Calibration Procedures
86
Keysight E3631A Service Guide
15
"CAL SETUP 5" now appears on the display (press "Calibrate"
key).
-25 supply
16
"V LO -0.5000 V" appears on the display (change the display to
match the DVM reading; then press "Calibrate" key).
V
-25 supply
17
"V HI -24.750 V" appears on the display (change the display to
match the DVM reading; then press "Calibrate" key).
V
-25 supply
18
"CAL SETUP 6" now appears on the display (press "Calibrate"
key).
-25 supply
19
"I LO 0.100 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
-25 supply
20
"I HI +0.950 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
-25 supply
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...