3
Calibration Procedures
72
Keysight E3631A Service Guide
Performance test record for Keysight E3631A
CV performance test record
Test Description
Actual Result
Specifications
Upper Limit
Lower Limit
CV Programming Accuracy @ 0 volts
(DVM reading)
+6V supply
+25V supply
-25V supply
+0.0050 V
+0.0200 V
+0.0200 V
-0.0050 V
-0.0200 V
-0.0200 V
CV Readback Accuracy @ 0 volts
+6V supply
+25V supply
-25V supply
DVM + 0.0050 V
DVM + 0.0100 V
-(DVM + 0.0100) V
DVM - 0.0050 V
DVM - 0.0100 V
-(DVM - 0.0100) V
CV Programming Accuracy @ Full Scale
(DVM reading)
+6V supply
+25V supply
-25V supply
6.0110 V
25.0325 V
-25.0325 V
5.9890 V
24.9675 V
-24.9675 V
CV Readback Accuracy @ Full Scale
+6V supply
+25V supply
-25V supply
DVM + 0.0110 V
DVM + 0.0225 V
-(DVM + 0.0225) V
DVM - 0.0110 V
DVM - 0.0225 V
-(DVM - 0.0225) V
CV Load Regulation
+6V supply
+25V supply
-25V supply
Maximum change:
<2.6 mV
<4.5 mV
<4.5 mV
CV Line Regulation
+6V supply
+25V supply
-25V supply
Maximum change:
<2.6 mV
<4.5 mV
<4.5 mV
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...