Calibration Procedures
3
Keysight E3631A Service Guide
67
6
Record the value displayed on the controller. This value should be within the
limits specified below for each output tested.
7
Program the selected output current to the full scale value (see Table 3-2) by
sending the commands:
CURR 5.0 for the +6V supply or
CURR 1.0 for the ±25V supply
8
Divide the voltage drop (DVM reading) across the current monitoring resistor
(RM) by its resistance to convert to amps and record this value (IO). This value
should be within the limits specified below for each output tested.
9
Readback the output current from the selected output over the remote
interface by sending the command:
MEAS:CURR? {P6V|P25V|N25V}
10
Record the value displayed on the controller. This value should be within the
limits specified below for each output tested.
Output
Readback Accuracy
+6V Output
I
O
± 10 mA
+25V Output
I
O
A ± 4 mA
-25V Output
I
O
A ± 4 mA
Output
Programming
Accuracy
+6V Output
5 A ± 20 mA
+25V Output
1 A ± 5.5 mA
-25V Output
1 A ± 5.5 mA
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...