Calibration Procedures
58
Keysight E3631A Service Guide
5
Readback the output voltage of the selected output over the remote interface
by sending the command:
MEAS:VOLT? {P6V|P25V|N25V}
6
Record the value displayed on the controller. This value should be within the
limits specified below for each output tested.
7
Program the selected output's voltage to full scale value by sending the
commands.
VOLT 6.0 for the +6V supply or
VOLT 25.0 for the ± 25V supply
8
Record the output voltage reading on the digital voltmeter (DVM). The
readings should be within the limits specified below for each output tested.
Output
Programming
Accuracy
+6V Output
0 ± 5 mV
+25V Output
0 ± 20 mV
-25V Output
0 ± 20 mV
Output
Readback Accuracy
+6V Output
DVM ± 5 mV
+25V Output
DVM ± 10 mV
-25V Output
-(DVM ± 10 mV)
Output
Programming
Accuracy
+6V Output
6 V ± 11 mV
+25V Output
25 V ± 32.5 mV
-25V Output
-(25 V ± 32.5 mV)
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...