Service
5
Keysight E3631A Service Guide
111
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when serving the power
supply or any electronic device.
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to dissipate static charge.
– Use a conductive wrist strap to dissipate static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastic, styrofoam, vinyl, paper, and other static-generating
materials from the immediate work area.
– Use only anti-static solder suckers.
Surface Mount Repair
Surface mount components should only be removed using soldering irons or
desoldering stations expressly designed for surface mount components. Use of
conventional solder removal equipment will almost always result in permanent
damage to the printed circuit board and will void your Keysight Technologies
factory warranty.
To Replace the Power-Line Fuse
The power-line fuse is located within the power supply's fuse-holder assembly on
the rear panel (see in Chapter 2). For 100 or 115 Vac operation, you must use a
2.5 AT slow-blow fuse (Keysight part number 2110-0913). For 230 Vac operation,
you must use a 2 AT slow-blow fuse (Keysight part number 2110-0587).
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...