3
Calibration Procedures
82
Keysight E3631A Service Guide
10
Select the first current calibration point for the +6V supply.
The display shows the first current calibration point for the +6V supply.
11
Read the DVM and change the first current value on the display to match the
computed current (DVM reading ? by shunt resistance).
For example, if the computed value is 0.099 A, adjust the current to 0.099 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
12
Pressing the "Calibrate" key saves the change and selects the second current
calibration point for the +6V supply.
If the entered number is within an acceptable range, an "ENTERED" message
appears for one second. The display now shows the second current calibration
point for the +6V supply. If the entered number is not correct, an "INVALID
DATA" message appears for one second and the display shows the first current
calibration point again.If the entered number is within an acceptable range, an
"ENTERED" message appears for one second. The display now shows the
second current calibration point for the +6V supply. If the entered number is
not correct, an "INVALID DATA" message appears for one second and the
display shows the first current calibration point again.
13
Read the DVM and change the second current value on the display to match
the computed current (DVM reading ? by shunt resistance). For example, if the
computed value is 4.999 A, adjust the current to 4.999 A using the knob and
arrow keys.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...