To check the rated voltages of the power supply
The following procedures check to ensure that the power supply develops its
rated voltage outputs with no load and properly responds to operation from the
front panel.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; all outputs are
disabled (the OFF annunciator turns on); the display is selected for the +6V
supply (the +6V annunciator turns on); and the knob is selected for voltage
control.
2 Enable the outputs.
The OFF annunciator turns off and the +6V and CV annunciators are lit. The
blinking digit can be adjusted by turning the knob. Notice that the display is in
the meter mode. "Meter mode" means that the display shows the actual
output voltage and current.
3 Check that the front-panel voltmeter properly responds to knob control for
+6V supply
.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero.
4 Ensure that the voltage can be adjusted from zero to the maximum rated
value.
1
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob
until the voltmeter indicates 6.0 volts.
5 Check the voltage function for the +25V supply.
Select the meter and adjust selection key for the +25V supply. The CV
annunciator is still lit and the +25V annunciator will turn on. Repeat steps (3)
and (4) to check the voltage function for the +25V supply.
6 Check the voltage function for the –25V supply.
Select the meter and adjust selection key for the –25V supply. The CV
annunciator is still lit and the –25V annunciator will turn on. Repeat steps (3)
and (4) to check the voltage function for the –25V supply.
1
You can use the resolution selection keys to move the blinking digit to the right
or left when setting the voltage.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...