Calibration Procedures
3
Keysight E3631A Service Guide
81
7
Pressing the "Calibrate" key saves the change and selects the second voltage
calibration point for the +6V supply.
If the entered number is within an acceptable range, an "ENTERED" message
appears for one second. The display now shows the second voltage calibration
point for the +6V supply. If the entered number is not correct, an "INVALID
DATA" message appears for one second and the display shows the first voltage
calibration point again.
8
Read the DVM and change the second voltage value on the display to match
the measured voltage.
For example, if the DVM reads 5.9990 V, adjust the voltage to 5.9990 V using
the knob and arrow keys.
9
Pressing the "Calibrate" key saves the new calibration constants for the +6V
output voltage and goes to the current calibration mode for the +6V supply.
If the entered number is within an acceptable range, a "CALIBRATING"
message appears for one second to indicate that the calibration is successful
and that new calibration constants of "SETUP 1" are stored. Then, the display
shows the above message to indicate that the power supply is ready for the
current calibration for +6V supply.
If the entered number is not correct, an "INVALID DATA" message appears for
one second and the display shows the second voltage calibration point again.
If the calibration fails, a "CAL FAIL" message appears for one second and the
display shows the "CAL SETUP 1" for the voltage calibration of the +6V supply
again.
*Connect an appropriate shunt (see
) across the +6V supply's output
terminals, and connect a digital voltmeter across the shunt resistor for the
current calibration.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...