Quick Start
2
Keysight E3631A Service Guide
37
7 Check the current function for the +25V supply.
Disable the outputs by pressing the “Output On/Off” key and connect a short
across (+) and (COM) output terminals of the +/–25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the +25V supply.
8 Check the current function for the –25V supply.
Disable the outputs by pressing the “Output On/Off” key and connect a short
across (-) and (COM) output terminals of the +/–25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the –25V supply.
1
You can use the resolution selection keys to move the blinking digit to the right or left when setting
the current.
NOTE
If an error has been detected during the output checkout procedures, the
ERROR annunciator will turn on. See "Error Messages" for more information,
starting on page 113 in chapter 5 of the User's Guide.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...