Calibration Procedures
3
Keysight E3631A Service Guide
57
Constant voltage (CV) verifications
Constant voltage test setup
If more than one meter or a meter and an oscilloscope are used, connect each to
the (+) and (-) terminals by a separate pair of leads to avoid mutual coupling
effects. Use coaxial cable or shielded 2-wire cable to avoid noise pick-up on the
test leads.
Voltage programming and readback accuracy
This test verifies that the voltage programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the remote
interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test to
avoid round off errors.
1
Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in
.
2
Turn on the power supply. Enable the outputs and select the desired output to
be tested by sending the commands:
OUTP ON?
INST {P6V|P25V|N25V}
3
Program the selected output to zero volts and maximum programmable
current (see
) by sending the commands:
4
Record the output voltage reading on the digital voltmeter (DVM). The
readings should be within the limits specified below for each output tested.
Also, note that the CV, Adrs, Lmt, and Rmt annunciators are on.
VOLT 0
CURR 5.15
for the +6V supply or
CURR 1.03
for the ±25V supply
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...