Calibration Procedures
3
Keysight E3631A Service Guide
83
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
14
Pressing the "Calibrate" key saves the new calibration constants for the +6V
output current and goes to the voltage calibration mode for the +25V supply.
If the entered number is within an acceptable range, a "CALIBRATING"
message appears for one second to indicate that the calibration is successful
and that new calibration constants of "SETUP 2" are stored. Then, the display
shows the above message to indicate that the power supply is ready for the
voltage calibration for +25V supply.
If the entered number is not correct, an "INVALID DATA" message appears for
one second and the display shows the second voltage calibration point again.
If the calibration fails, a "CAL FAIL" message appears for one second and the
display shows the "CAL SETUP 2" for the current calibration of the +6V supply
again.
15
Repeat steps (5) through (14) for the voltage and current calibration of the
+25V supply (r6V with +25V in all the text).
You can calibrate the output voltage and current of the +25V supply easily in
the same manner. Notice that you should connect a shunt across (+) and
(COM) terminals of the ±25V supply for the +25V supply current calibration.
16
Repeat steps (5) through (14) for the voltage and current calibration of the
-25V supply (r6V with -25V in all the text).
You can also calibrate the output voltage and current of the -25V supply easily
in the same manner. Notice that you should connect a shunt across (-) and
(COM) terminals of the ±25V supply for the -25V supply current calibration.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...