5
Service
108
Keysight E3631A Service Guide
Operating Checklist
Before returning your power supply to Keysight Technologies for service or repair
check the following items:
Is the Power Supply Inoperative?
– Verify that the ac power cord is connected to the power supply.
– Verify that the front-panel power switch is depressed.
– Verify that the power-line fuse is installed:
– Use the 2.5 AT, 250 V fuse for 100 or 115 Vac operation.
– Use the 2 AT, 250 V fuse for 230 Vac operation.
– Verify the power-line voltage setting.
– See "To prepare the power supply for use" in Chapter 2.
Does the Power Supply Fail Self-Test?
– Verify that the correct power-line voltage is selected.
See "To prepare the power supply for use" in Chapter 2.
– Remove all load connections to the power supply.
Ensure that all terminal connections are removed while the self-test is
performed.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...