Troubleshooting Hints
This section provides a brief check list of common failures. Before troubleshooting
or repairing the power supply, make sure that the failure is in the instrument
rather than any external connections. Also make sure that the instrument is
accurately calibrated. The power supply's circuits allow troubleshooting and
repair with basic equipment such as a digit multimeter and a 100 MHz
oscilloscope.
Unit is Inoperative
– Verify that the ac power cord is connected to the power supply.
– Verify that the front-panel power switch is depressed.
– Verify that the power-line fuse is installed:
– Use the 2.5 AT, 250 V fuse for 100 or 115 Vac operation..
– Use the 2 AT, 250 V fuse for 230 Vac operation..
– Verify the power-line voltage setting.
– See
To prepare the power supply for use
Unit Reports Errors 740 to 748
These errors may be produced if you accidentally turn off power the unit during a
calibration or while changing a non-volatile state of the instrument. Recalibration
or resetting the state should clear the error. If the error persists, a hardware failure
may have occurred.
Unit Fails Self-Test
Verify that the correct power-line voltage setting is selected. Also, ensure that all
terminal connections are removed while the self-test is performed. Failure of the
DAC U12
[1]
on the top board will cause many self-test failures.
[1]
For serial MY53xx6xxx, failure of the DAC U13 on the top board will cause many
self-test failures.
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...