To set up the screen saver
You can configure the instrument to turn on a screen saver after a specified idle time.
1.
Press
[Utility] > User Options > Preferences > Screen Saver
.
2.
Press
Screen Saver
and choose whether you want the screen save to be off (disabled), a
blank screen, a Keysight logo, or custom text (press the
Text
softkey that appears to specify
that text.)
3.
Press
Wait
to select the idle time before the screen saver activates. The default is 120
minutes.
4.
Press
Preview
to see the screen saver. When finished, press any key or turn any knob.
To set the instrument's clock
You can set the instrument's real-time clock for the date/time stamp that appears on printouts
and file attributes. The clock uses a 24-hour format, which means that 1:00 PM is 13:00.
1.
Press
[Utility] > User Options > Date/Time
.
2.
Press the
Year
,
Month
,
Day
,
Hour
or
Minute
softkey. Then rotate the
Entry
knob to set to
the desired number. Repeat as necessary until the date and time are correct.
If a day is selected and the month or year is changed so the day is invalid, the day is automatically
adjusted.
To Use the Rear-Panel Reference Clock
To use an external clock signal on the rear-panel Ref In connector (10 MHz, 0 to 3.3 V) instead
of the instrument's internal clock, press
[Utility] > User Options > Rear Panel > Ref In Clock
and choose
Use
. The default is
Ignore
.
Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide 85