External Transducer Input
The Measurement Limit from the center conductor to the outer conductor is 10 VAC rms
or VDC. The Measurement Limit from center conductor or the outer conductor to the
chassis is 1000 VAC rms or VDC.
Before energizing any measurement circuits, ensure that the external transducer’s BNC
terminal is properly locked onto the instrument’s BNC receptacle and that all meas-
urement circuits are properly configured. Never modify any connections to the instru-
ment or touch any input terminals while measurement circuits are energized.
Always use a cable that is suitably rated and insulated for the maximum voltage that
could be applied to any terminal. Be sure to consider both the differential and common
mode voltages that may appear on your measurement circuit when choosing an appro-
priate cable.
Always allow adequate time for circuits being measured to fully de-energize after remov-
ing power to these circuits. Do not rely on the reported readings from the instrument to
determine whether a circuit is de-energized.
16
Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide