4- 14
Keysight B1507A User’s Guide, Edition 1
GUI Reference
Capacitance Measurement Software
Creates new settings to measure the Cp-G model capacitance between the
collector/drain and the emitter/source terminals for the swept collector-emitter
or drain-source voltage.
•
Base/Gate Voltage Sweep
Creates new settings to measure the Cs-Rs model capacitance between the
base/gate and the emitter/source terminals for the swept base-emitter or
gate-source voltage.
General settings
In this area, you can set the following measurement conditions.
•
Frequency
Specify the output frequency of the AC signal from MFCMU.
Setting range: 1 kHz to 1 MHz
•
AC Level
Specify the output level of the AC signal from MFCMU.
Setting range: 0 to 250 mV
•
N PLC
Specify the measurement time (integral time) in the number of power line cycle.
Setting range: 1 to 100
•
Phase Compensation
Specify one from the following phase compensation modes.
•
Auto
Uses the predefined phase compensation data.
•
Adaptive
Compensates the phase at each measurement point before the measurement.
It is useful when you cannot measure in the Auto compensation mode by
MFCMU oscillation.
The following parameter can be shown or hidden by clicking the chevron
of the
Details.
•
AC Level
•
N PLC
•
Phase Compensation