Keysight B1507A User’s Guide, Edition 1
3- 13
How To Perform Measurement
How To Perform I/V Measurement
Select the step value immediately after where you want to insert new step value
and click the Insert button
.
•
To change an output order of step values
Click the Move Down and the Move Up buttons
to move down or up an
order of the selected step value.
•
To delete a step value
Click the Delete button
to delete the selected step value.
By clicking the
Details
chevron, the area expands to set more detail conditions.
For more information on setting items, see “Setup” on page 4-38.
Step 4.
Enter or select the setup conditions for the drain of MOSFET in Drain/Collector
Voltage Pulse Sweep at the right side of Setup area.
Select or specify a sweep mode, a sweep start value, a sweep stop value, and a
voltage compliance.
•
Sweep Mode
Select a sweep mode from the Sweep Mode drop-down list.
LinearSingle
One-way linear sweep from the Start (sweep start value) to
the Stop (sweep stop value).
LogSingle
One-way logarithmic sweep from the Start (sweep start
value) to the Stop (sweep stop value).
LinearDouble
Linear-sweep from the Start (sweep start value) to the Stop
(sweep stop value) and then return linear-sweep from the
Stop to the Start.