Keysight B1507A User’s Guide, Edition 1
1- 29
Introduction
Measurement Resources
Table 1-8
HVSMU Voltage Ranges and Resolutions
Table 1-9
HVSMU Current Ranges and Resolutions
Range
Output Value and
Measurement Value
Setting/
Measurement
Resolution
Maximum Current
|Imax|
200 V
0
|V|
200 V
200
V
8 mA
500 V
0
|V|
500 V
500
V
8 mA
1500 V
0
|V|
1500 V
1.5 mV
8 mA
3000 V
0
|V|
3000 V
3 mV
4 mA
Range
Output Value and
Measurement Value
Setting/
Measurement
Resolution
Maximum Voltage
|Vmax|
1 nA
a
a. DC only
0
|I|
1.15 nA
10 fA
3000 V
10 nA
0
|I|
11.5 nA
100 fA
3000 V
100 nA
0
|I|
115 nA
100 fA
3000 V
1
A
0
|I|
1.15
A
1 pA
3000 V
10
A
0
|I|
11.5
A
10 pA
3000 V
100
A
0
|I|
115
A
100 pA
3000 V
1 mA
0
|I|
1.15 mA
1 nA
3000 V
10 mA
0
|I|
8 mA
10 nA
1500 V
0
|I|
4 mA
3000 V
Summary of Contents for B1507A
Page 1: ...Keysight Technologies B1507A Power Device Capacitance Analyzer User s Guide...
Page 27: ...1 Introduction...
Page 59: ...2 Installation...
Page 79: ...3 How To Perform Measurement...
Page 109: ...4 GUI Reference...
Page 189: ......