4- 56
Keysight B1507A User’s Guide, Edition 1
GUI Reference
I/V Measurement Software
NOTE
If DUT is a diode, connect the cathode terminal to 2. Collector/Drain terminal, and
the anode terminal to 3. Emitter/Source terminal.
Base/Gate
Resistor
If the MCSMU is specified to the Base/Gate terminal as the measurement resource,
you can select one from the following base/gate resistances.
•
0
•
10
•
100
•
1 k
The base/gate resistance is effective to restrain the oscillation of DUT that occurs
during the ultra high current measurement and so on.
Collector/Drain
Resistor
If the MVSMU is specified to the Collector/Drain terminal as the measurement
resource, you can select one from the following collector/drain resistances.
•
0
•
100 k
The collector/drain resistance is effective to restrain the over current that occurs
during the breakdown measurement and so on.