Keysight B1507A User’s Guide, Edition 1
2- 5
Installation
Requirements
Installation Requirements
WARNING
Do not operate the instrument in dusty environment, or in the presence of
flammable gasses, corrosive gasses, or fumes.
Ne pas utiliser l’appareil dans un endroit poussiéreux, ou en présence de gaz
inflammables, corrosifs ou de fumée.
•
Environmental conditions for this instrument are documented in the previous
page. In principle, this instrument should only be operated indoors in a
controlled environment.
•
Install the instrument horizontal place then face the roof up.
•
The dimensions of this instrument are shown below. Fan cools the instrument by
drawing air through the sides and exhausting it out the back. The instrument
must be installed in a location that allows sufficient space at the sides and back
of the instrument for adequate air circulation.
•
Mainframe: 420 mm (W)
575 mm (D)
330 mm (H)
•
Selector: 420 mm (W)
575 mm (D)
75 mm (H)
•
Test fixture: 340 mm (W)
345 mm (D)
200 mm (H)
If the airflow is restricted, the internal operating temperature will be higher. This
may reduce the instrument’s reliability, or cause the thermal-protection circuits
to turn the instrument off.
•
Make sure that there is enough space for connecting/disconnecting the power
cable easily.