Keysight B1507A User’s Guide, Edition 1
4- 41
GUI Reference
I/V Measurement Software
Specify the hold time of the sweep start voltage. At the sweep start point, the
start voltage is forced for more than the specified time.
Setting range: 0 to 655.35 s
For pulse output measurement:
•
Auto Period
Enable or disable the automatic adjustment for the pulse period. To enable this,
mark the check box.
•
Pulse Period
Specify the pulse period.
Setting range: 5 ms to 5 s
If the Auto Period is enabled, this is automatically adjusted to minimum period
corresponding to the pulse width.
•
Aperture
Specify the measurement time at each pulse.
Setting range: 2
s to 20 ms
•
Hold Time
Specify the hold time of the sweep start voltage. At the sweep start point, the
start voltage is forced for more than the specified time.
Setting range: 0 to 655.35 s
The following parameter can be shown or hidden by clicking the chevron
of the
Details.
•
Aperture
•
Hold Time