1-4
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Section 1: Getting Started
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Figure 1-1
Model 2308 front panel
Figure 1-2
Model 2308 rear panel
Make sure that the maximum combined channel output is not exceeded (see complete
Specifications at
www.keithley.com
). Also, do not exceed 3 A when using the power supply as a
sink. For output voltages exceeding 5 V, the maximum sink current is less than 3 A (derate the
maximum sink current 0.2 A for each volt over 5 V).
NOTE
The output from each channel is isolated from the other channel.
CAUTION
Allow for adequate ventilation around the front and rear-panel openings to
prevent the unit from overheating which may degrade performance.
NOTE
When using the power supply as a sink (negative polarity), the power supply is
dissipating rather than sourcing power (see
Sink operation
).
POWER
DISPLAY
OPERATE
ENTER
SET
MENU
LOCAL
2308 Portable Device Battery/Charger Simulator
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+ + + _ _ _
SOURCE
SENSE
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX MAX.
RELAY
CONTROL
24VDC MAX
DVM IN
-30 VDC MAX
ANALOG
OUT
mA
(5/50)
A
(0.5/5)
+ + + _ _ _ _ +
SOURCE
SENSE
DVM IN
SOURCE
OUTPUT #2
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...