Section 13: Calibration
Figure 13-2
Connections for 5 A, 500 mA, and 50 mA current calibration
9. At this point, the unit will prompt for 5 A Channel #1 full-scale calibration output:
CAL 5 A CHAN 1
SET 1.90000 A
10. Using the edit keys, adjust the set value to 1.9000 A, then press
ENTER
. The unit will
prompt you for the DMM reading, which calibrates the 5 A current limit for Channel #1:
CAL 5 A CHAN 1
READ1 1.90000 A
11. Note the DMM voltage reading, then calculate the current from that reading and the actual
4
Ω
resistance value: I = V/R. Adjust the Model 2308 current display value to agree with the
calculated current value, and press
ENTER
.
12. The Model 2308 will prompt for another DMM reading, which is used for 5 A measurement
calibration of Channel #1:
CAL 5 A CHAN 1
READ2 1.90000 A
13. Again, calculate the current from the new DMM reading and 4
Ω
resistor value. Adjust the
Model 2308 current display reading to agree with the new current, then press
ENTER
.
14. The unit will now prompt you for Channel #1 500 mA full-scale calibration:
CAL 500mA CHAN 1
ALL READY TO DO?
15.
Press
ENTER
.
16. The unit will then prompt you for the DMM reading:
CAL 500mA CHAN 1
READ1 450.000 mA
17. Note the DMM voltage reading, then calculate the current from that reading and the actual
4
Ω
resistance value: I = V/R. Adjust the Model 2308 current display value to agree with the
calculated current value, and press
ENTER
.
18. The unit will prompt for Channel #1 50 mA full-scale calibration:
CAL 50mA CHAN 1
ALL READY TO DO?
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+
+
+ _
_
_
SOURCE
SENSE
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX MAX.
RELAY
CONTROL
24VDC MAX
DVM IN
-30 VDC MAX
ANALOG
OUT
mA
(5/50)
A
(0.5/5)
+
+
+ _
_
_
_
+
SOURCE
SENSE
DVM IN
SOURCE
OUTPUT #2
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
ý2
ý4
FREQ
TEMP
REL
TRIG
INFO LOCAL
EXITENT
ER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
ý 4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2308
Input LO
Input HI
Source -
4
W
Resistor
Note: Use 4-wire connections
to resistor terminals.
Sense -
Sense +
Use OUTPUT #1 connections shown and used
for 5 A, 500 mA, and 50 mA. Use OUTPUT #2
for Charger Channel (5 A only).
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...