B-4
Return to
Section Topics
2308-900-01 Rev. A / July 2008
Appendix B: Calibration Reference
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
:COUNT?
(:CALibration:PROTected:COUNt?)
Purpose
To request the number of times the Model 2308 has been calibrated.
Format
:cal:prot:count?
Response
Number of times calibrated.
Description
The
:COUNT?
query may be used to determine the total number of times the
Model 2308 has been calibrated. The calibration count will also be displayed
during the front-panel calibration procedure.
Example
:CAL:PROT:COUNT?
Request calibration count.
:DATE
(:CALibration:PROTected:DATE)
Purpose
To program the calibration date.
Format
:cal:prot:date <yyyy>, <mm>, <dd>
Parameter
<yyyy> = 2008 to 2107
<mm> = 1 to 12
<dd> = 1 to 31
Query
:cal:prot:date?
Response
<yyyy> , <mm> , <dd>
Description
The :DATE command allows you to store the calibration date in instrument
EEROM for future reference. You can read back the date from the instrument by
using the :DATE? query. The calibration date will also be displayed during the
front-panel calibration procedure.
Example
:CAL:PROT:DATE 2008,3,1
Send cal date (3/1/2008).
:INIT
(:CALibration:PROTected:INIT)
Purpose
To initiate calibration.
Format
:cal:prot:init
Description
The
:INIT
command initiates the calibration process and must be sent before all
other commands except
:CODE
.
Example
:CAL:PROT:INIT
Initiate calibration.
NOTE
The year, month, and day parameters must be delimited by commas.
NOTE
The :INIT command should be sent only once at the beginning of the calibration
procedure. Do not send
:INIT
before each calibration step.
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...