Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 13: Calibration
Step 3: Perform charger channel calibration steps
Figure 13-4
Charger channel connections for voltage calibration
1. Connect the OUTPUT #2 SOURCE, SENSE, and DVM IN terminals to the DMM as shown
in
Figure 13-4
.
DMM input connections:
HI: , SENSE +, and DVM IN +
LO: SOURCE -, SENSE -, and DVM IN -
2. The Model 2308 will prompt you to set the full-scale Channel #2 output voltage:
CAL VOLTS CHAN 2
SET 14.0000 V
3. Use the edit keys to set the voltage to 14.0000V, then press
ENTER
.
4. The unit will prompt you for the DMM reading, which will be used to calibrate the Channel #2
full-scale output voltage:
CAL VOLTS CHAN 2
READ1 14.0000 V
5. Using the edit keys, adjust the Model 2308 voltage display to agree with the DMM voltage
reading, then press
ENTER
. The unit will prompt for another DMM reading, which will be
used to calibrate the full-scale Channel #1 measurement function:
CAL VOLTS CHAN 2
READ2 14.0000 V
6. Using the edit keys, adjust the display to agree with the new DMM voltage reading, then
press
ENTER
. The unit will prompt for Channel #2 DVM full-scale calibration:
CAL DVM CHAN 2
ALL READY TO DO?
7. Press
ENTER
to complete Channel #2 DVM full-scale calibration.
8. Connect the digital multimeter volts input and characterized 4
Ω
resistor to the Model 2308
OUTPUT #2 SOURCE terminals using the same general connections shown in
Figure 13-2
.
Be sure to observe proper polarity ( to DMM INPUT HI; SOURCE - to INPUT
LO).
9. Be sure the digital multimeter DC volts function and auto-ranging are still selected.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+
+
+ _
_ _
SOURCE
SENSE
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX MAX.
RELAY
CONTROL
24VDC MAX
DVM IN
-30 VDC MAX
ANALOG
OUT
mA
(5/50)
A
(0.5/5)
+
+
+ _
_ _
_
+
SOURCE
SENSE
DVM IN
SOURCE
OUTPUT #2
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
ý2
ý4
FREQ
TEMP
REL
TRIG
INFO LOCAL
EXITENT
ER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
ý 4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2308
Input LO
Input HI
Source -
DVM IN -
DVM IN +
Sense +
Sense -
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...