Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 11: SCPI Tables
:FAST <b>
Enable or disable pulse current fast readings.
OFF
:FAST?
Query pulse current fast reading setting.
:SEARch <b>
Enable or disable pulse current search.
ON
:SEARch?
Query pulse current search setting.
:DETect <b>
Enable or disable pulse current detection mode.
OFF
:DETect?
Query pulse current detection mode setting.
:TimeOUT <NRf>
Specify length of timeout: 5ms - 32s incrementing
in 1ms.
1 (sec)
:TimeOUT?
Query pulse current timeout setting.
:LINTegration
Path to configure long integration measurements
(refer to
Section 4
):
:TEDGe <name>
Set trigger edge to start long integration
measurement (RISING, FALLING, NEITHER).
RISING
:TEDGe?
Query trigger edge setting.
:TIME <NRf>
Set integration time in seconds: X to 60 where X
is 0.850 for 60Hz line frequency or 0.840 for 50Hz
line frequency.
1.00
:TIME?
Query integration time.
:AUTO
Power supply sets integration time.
:TLEVel
Path to set trigger level.
[:AMP] <NRf>
Set the trigger level in amps for use on the 5 A
current range: 0.0 A-5.0 A in 5 mA steps.
0
[:AMP]?
Query setting for trigger level in amps for use on
the 5 A current range.
:HUNDred <NRf>
Set the trigger level in amps for use on the 500
mA current range: 0.0 mA-500 mA in 0.5 mA or
500
μ
A steps.
0
:HUNDred?
Query setting for trigger level in amps for use on
the 500 mA current range.
:FIFTy <NRf>
Set the trigger level in amps for use on the 50 mA
current range: 0.0 mA-50 mA in 0.05 mA or 50 uA
steps
0
:FIFTy?
Query setting for trigger level in amps for use on
the 50 mA current range.
:FIVE <NRf>
Set the trigger level in amps for use on the 5 mA
current range: 0.0 mA-5 mA in 0.005 mA or 5
μ
A
steps.
0
:FIVE?
Query setting for trigger level in amps for use on
the 5 mA current range trigger level setting.
:TimeOUT <NRf>
Specify length of timeout: 1 to 63 (seconds).
16
Table 11-4
SENSe command summary (refer to
Table 2-3
,
Table 3-2
, and
Table 4-2
)
Command
Description
Default
parameter SCPI
*
No query or short form exists.
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...