2308-900-01 Rev. A / July 2008
Return to
Section Topics
4-11
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 4: Long Integration Measurements
SCPI programming
The commands for long integration measurements are summarized in
Table 4-2
(a listing following
the table contains specific command notes).
Programming examples
demonstrate how to use
these commands.
NOTE
Setting the trigger level and/or the current range may cause LONG INT TRIG
NOT DETECTED to appear.
Table 4-2
SCPI commands - long integration measurements
Command
Description
Default
:SENSe[1]
SENSe subsystem for battery channel (#1):
:FUNCtion “LINTegration”
Select long integration measurement function.
VOLT
:LINTegration
Long integration configuration:
:TIME <NRf>
Set integration time (in sec): X to 60 (where X
is 0.850 for 60Hz, or 0.840 for 50Hz).
1
:AUTO
Integration time set automatically.
:TLEVel <NRf>
Path to set trigger level feature:
[:AMP] <NRf>
Set the trigger level in amps for use when on
the 5A current range: 0.0 A-5.0 A in 5 mA steps
0.0
:HUNDred <NRf>
Set the trigger level in amps for use when
on the 500mA current range: 0.0 mA-500 mA in
0.5 mA or 500
μ
A steps.
0.0
:FIFTy <NRf>
Set the trigger level in amps for use when on
the 50mA current range: 0.0 mA-50 mA in 0.05 mA
or 50
μ
A steps.
0.0
:FIVE <NRf>
Set the trigger level in amps for use when
on the 5 mA current range: 0.0 mA-5 mA in
0.005 mA or 5
μ
A steps.
0.0
:TEDGe <name>
Select trigger edge to initiate the
measurement:
RISING, FALLING, or NEITHER.
RISING
:TimeOUT <NRf>
Specify length of timeout: 1–63 (sec).
16
:SEARch <b>
Enable or disable pulse search.
ON
:FAST <b>
Enable or disable long integrations fast
readings.
OFF
:DETect <b>
Enable or disable pulse detection mode.
OFF
:SENSe2
SENSe subsystem for the charger channel (#2):
:FUNCtion “LINTegration”
Select long integration measurement function.
VOLT
:LINTegration
Long integration configuration:
:TIME <NRf>
Set integration time (in sec): X to 60 (where X
is 0.850 for 60Hz, or 0.840 for 50Hz).
1
:AUTO
Integration time set automatically.
:TLEVel <NRf>
Set trigger level in amps: 0–5 (5mA resolution). 0.0
Summary of Contents for 2308